Radiancy FSD+ Bedienungsanleitung Seite 104

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 109
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 103
Lifvation
http://3dLipoMed.Lifvation.com
Cryo Maintenance
April 15, 2014
LIFVATION, http://www.lifvation.com, all
rights reserved.
104
# What you
are
checking:
Procedure Frequency
y1 Water
Level
Always ensure Water
tank is full and follow
the instructions on the
water filling document.
Weekly
y2 Plate
cooling
efficiency
Check that both
cooling plates in the
hand sets are chilling
down
Daily
y3 Vacuum
efficiency
Always ensure the filter
on the back of the
machine is free of fluid.
Empty immediately if
signs of fluid are
present ensuring that
the O ring remains in
place to prevent loss of
vacuum.
Daily
y4 Replace Cryo head
filter at least once per
month or sooner if
experiencing lack of
vacuum.
Monthly
y5 Vacuum
Safety
Always check both
emergency stop
buttons are functioning
prior to treatment.
Daily